|
|
TM 3-6665-225-12
Section III. OPERATION UNDER UNUSUAL CONDITIONS
CLOSED position (fig. 2-5, A). Unscrew plug (1).
2-14. General
This section contains instructions for operating the
it fully clockwise and release it. Rotate HORN VOL-
equipment in tropical environments, in tempera-
BATTERY TEST knob fully clockwise.
tures below 40F., in blowing sand or dust, and in
i. Notify personnel within audible range that
raining, sleeting, or snowing environments.
alarm system is to be tested.
NOTE
NOTE
In tropical environments, inspect the M42
If any startup tests fail, refer to table 3-3
alarm unit batteries every 15 days for elec-
for operator's troubleshooting information.
trolyte leakage. Replace defective batteries
j. Startup M43 detector unit as follows:
immediately. Replace batteries installed in
NOTE
M42 alarm units every 30 days as a mini-
In temperature below 40F., M43 detector
mum.
unit may require warmup to start pump
NOTE
assembly motor. Lower temperatures re-
If the alarm system is to be operational for
quire longer warmup, but no longer than
two or more continuous missions, perform
50 minutes.
checks and reservicing procedures (para 2-
(1) Connect power source connector to 24 VDC
10) after each 12-hour mission.
INPUT connector (11, fig. 1-2). If M10 power supply
is used, set power switch (3, fig. 2-2) to correct input
2-15. Operation in Temperatures Below
voltage position.
40F.
(2) If M43 detector unit horn (2, fig. 1-2) sounds
or pump assembly motor runs, proceed to (4) below.
graph 2-6.
(3) If horn does not sound or pump assembly
does not run, M43 detector unit is in warmup mode.
sensitivity check bottle (6), and one reservoir as-
Do not proceed until M43 detector unit horn sounds
sembly (5) from M229 refill kit. Thaw liquid in sen-
or pump assembly motor runs. Then proceed to (4)
sitivity check bottle and in reservoir assembly if fro-
below.
zen.
(4) Position M42 alarm unit selector switch (8,
NOTE
from M43 detector unit REMOTE binding post (17,
When operating between 40F. and 20F.,
use either BA3517/U battery (7, fig. 1-1) or
from 24 VDC INPUT connector (11).
BB501/U battery (9, fig. 2-7) and M168 ca-
(5) Prepare solution as described in figure 2-3.
ble (5) as prime power source or standby
NOTE
battery for M10 power supply. When oper-
To prevent the M43 detector unit from
ating below 20F., use only BB501/U bat-
cooling off, keep bottom case assembly
tery and M168 cable as primary power
separated from detector assembly for as
source or standby battery for M10 power
short a time as possible.
supply. Use table 1-2 as a guide for deter-
(6) Release four clamping catches (12, fig. 1-2)
mining maximum number of missions be-
from bottom catches (13) and remove the detector
fore exchanging BA3517/U battery for a
assembly (6) from the bottom case assembly (9).
fresh battery or exchanging BB501/U
WARNING
battery for a freshly charged battery.
Solution in the reservoir assembly is caus-
d. Install alarm system in accordance with ap-
tic. Avoid contact with eyes and mouth.
propriate instructions in paragraphs 3-1 through
Flush from skin with water.
3-4.
CAUTION
NOTE
Keep weight assembly free from dirt after
Do not connect power to M43 detector unit
removing the reservoir assembly.
24 VDC INPUT connector until instructed
(7) Hold the detector assembly upright by its
to do so.
handle (2, fig. 1-3) and unscrew the reservoir as-
sembly (12). Allow the weight assembly to hang
AIR OUTLET port.
down. Discard used reservoir assembly in accor-
f. Remove electrical connector cover (10) from 24
dance with standing operating procedures.
VDC INPUT connector (11).
(8) Remove and discard cap from the pre-
g. Check that AIR INLET assembly is in
2-11
Change 2
|
Privacy Statement - Press Release - Copyright Information. - Contact Us |